Designated for use with BSL Class I and II organisms, such as E.coli. 70% of air is recirculated within the hood, 30% of the air is exhausted through a HEPA filter.
Class II B2
Protects users, products and environments from hazardous agents that require Biosafety Level 1, 2 or 3 containment. Dedicated exhaust system and remote blower discharge 100% of filtered exhaust air. Suitable for use with hazardous chemicals and fumes.
Class II C1
Toggle between A2 and B2 configurations. Protects personnel, product and lab environment by maintaining complete containment in event of exhaust system failure. For Biosafety Level (BSL) 1, 2, and 3 containment or BSL 4 with PPE
Purifier Logic+ discharges 100% of filtered exhaust air from work area via a dedicated exhaust system and remote blower; models need to be hard ducted to the outside